Writing a clinical evaluation report (CER) under the EU MDR works best as a chain of connected decisions, not a late documentation task. Manufacturers plan the evaluation early, run a defensible literature search, establish the state of the art, close data gaps through post-market clinical follow-up (PMCF), and keep the CER current with post-market data.
Many teams still treat the clinical evaluation as one more file in the technical documentation. It gets written last, signed off, and rarely connected back to the design, the risk management file, or the instructions for use (IFU). That wastes most of what the evaluation produces, because a CER written alongside those documents can change them.
The practices below apply to legacy devices, meaning devices previously CE marked under the MDD or AIMDD, and to devices that have never carried a CE mark. For legacy devices, MDCG 2020-6 describes the clinical evidence needed under the MDR.
When should you start writing the clinical evaluation report?
A manufacturer should start the clinical evaluation early in the device lifecycle, in parallel with the design, risk management, instructions for use, and post-market surveillance documents. Starting early lets the clinical evaluation shape those documents rather than merely describe them, and it exposes clinical data gaps while there is still time to close them.
MDR Article 61 and Part A of Annex XIV require manufacturers to plan, conduct, and document the clinical evaluation, so the real starting point is the clinical evaluation plan, not the report. Drafting that plan while the design is still open means an evidence gap can still drive a design decision. Found after the design is frozen, the same gap can mean generating new clinical data on a fixed device.
How do you run a literature search for a clinical evaluation?
A literature search for a clinical evaluation is a systematic, documented search of published data on the subject device, similar devices, alternative treatments, and the clinical background. Its protocol fixes the objectives, search strings, inclusion and exclusion criteria, and appraisal criteria in advance, so the search can be reproduced and defended in front of a Notified Body.
The literature search is one of the foundations of the CER and among its most time-consuming parts, which is the argument for doing it first. A well-designed protocol keeps irrelevant results to a manageable volume and raises the value of the data that remain. Run early, the search can also change the evaluation strategy. It can surface:
- Published data on your own device that you did not know existed, a finding that your device is standard of care among the alternatives, or an equivalent device with extensive clinical data.
- Established endpoints and expected acceptance criteria for a planned clinical investigation, taken from data on similar devices.
- Input for risk management and the IFU: clinical risks or adverse events nobody had considered, confirmation that known risks are well mitigated in practice, realistic risk rates, and evidence on whether each IFU warning or precaution is justified.
How does state of the art fit into a clinical evaluation report?
State of the art in a clinical evaluation is current, generally accepted clinical practice for the condition a device addresses, including available alternative treatments and the safety and performance they achieve. The EU Medical Device Regulation (MDR) uses this benchmark to judge whether a device’s benefit-risk ratio is acceptable, so the literature search must establish it.
State of the art does not mean the newest or most advanced technology. It means what clinicians currently accept as good practice, and the benchmark comes from the alternatives, not from the subject device. A search protocol built only around the subject device cannot establish it. Our guide to state of the art in practice covers how to define it.
How should you plan for clinical data gaps?
A clinical data gap is a claim about a device’s safety, performance, or clinical benefit that the available clinical evidence does not yet support. Manufacturers should document each gap in the clinical evaluation and state how it will be closed, through a post-market clinical follow-up (PMCF) study, a survey, or real-world evidence collection, and by when.
Finding gaps is a normal outcome of the evaluation. A CER that names a gap and links it to a PMCF activity with a timeline shows an assessor that the manufacturer understands its own evidence. For implantable and class III devices the first question is different: MDR Article 61(4) requires clinical investigations unless specific exceptions apply, so gap planning starts with whether an exception is available. Our guide to PMCF under the MDR explains the options for closing gaps after launch.
What do Notified Bodies look for in a clinical evaluation report?
Notified Bodies assess a clinical evaluation report against MDR Article 61 and Annex XIV, and they record the outcome in a clinical evaluation assessment report (CEAR) built on the MDCG 2020-13 template. The assessment follows the evaluation process, from the plan and literature appraisal to the benefit-risk conclusion and the post-market clinical follow-up plan, not only the finished report.
The template is public. MDCG 2020-13 is a Notified Body document, not a manufacturer deliverable, but reading it before you start drafting shows how an assessor moves through the evaluation, including whether the post-market surveillance plan and the PMCF plan are adequate.
How does a clinical evaluation report stay current after launch?
Under the EU MDR, the clinical evaluation and its documentation, including the clinical evaluation report (CER), must be updated throughout the device lifecycle with data from post-market clinical follow-up (PMCF) and post-market surveillance. For class III and implantable devices, the PMCF evaluation report, and the summary of safety and clinical performance where indicated, must be updated at least annually.
Treat the CER as a document with a change history, not a file with a submission date. Link it directly to the post-market surveillance and PMCF activities that feed it. Many of those activities overlap with the CER update, such as refreshing the literature search or reviewing vigilance data, so run separately they duplicate effort. The PMCF plan itself follows the MDCG 2020-7 template.
How can a clinical evaluation report support decisions beyond regulatory affairs?
A clinical evaluation report (CER) supplies evidence that other departments can use. Clinical teams draw on its endpoints and data gaps to design studies, marketing teams use its data to keep claims realistic and supported, and business development teams use its comparison with similar devices and alternatives to position the product.
A claim such as faster recovery either appears in the appraised data or it does not, and the CER settles that before the brochure is printed. The same comparison tells business development where the device sits against similar products.
How can Qserve support your clinical evaluation?
Qserve’s clinical and regulatory experts support manufacturers with drafting clinical evaluation reports (CERs), systematic literature searches and appraisals, strategic guidance on closing clinical evidence gaps, and review of draft CERs before Notified Body submission. Manufacturers can contact Qserve to scope the support their team needs.