In Qserve's review of the CERs it has supported, most clinical evaluations cleared Notified Body review in two rounds of questions, and one-round outcomes were the exception. What separates the two is rarely the volume of data. It is whether the evidence covers every variant and accessory in scope, and whether claims such as device lifetime stay within what the data supports.
Below is what sufficient clinical evidence means under EU MDR, how a Notified Body assesses it, what current EU data says about refusals and delays, and the round-by-round figures from Qserve's own files.
What does "sufficient clinical evidence" mean under EU MDR?
Under EU MDR, sufficient clinical evidence means clinical data and clinical evaluation results of enough quantity and quality to support a qualified assessment that a device is safe and achieves its intended clinical benefits when used as intended. Article 61 of the MDR requires the manufacturer to specify and justify the level of evidence each device needs.
The regulation sets no fixed quantity. The level of evidence scales with the device's characteristics, risk class, intended purpose, and the clinical claims made for it, which is why two devices in the same class can need very different clinical packages. For devices moving from the directives, MDCG 2020-6 gives the working definition and a suggested hierarchy of clinical evidence, and makes the uncomfortable point plainly: data that supported a CE mark under the MDD does not automatically meet the MDR standard.
How does a Notified Body assess a clinical evaluation report?
A Notified Body assesses a clinical evaluation report by checking whether the clinical evidence supports every clinical claim, indication, and device variant in the technical documentation. The clinical reviewer records the conclusions in a clinical evaluation assessment report (CEAR), using the template in MDCG 2020-13, including verdicts on equivalence, the benefit-risk determination, and post-market clinical follow-up.
In practice, the CEAR template is the reviewer's checklist. A CER organized so the reviewer can complete each CEAR section without searching through the technical file shortens the review, and fewer gaps turn into written questions. Endpoints, equivalence, and the link between residual risks and PMCF objectives are where reviewers dig deepest.
Why do MDR applications get refused or delayed?
The EU's 20th notified body survey (data to February 2026) shows the two main reasons notified bodies refuse MDR applications: the application falls outside the notified body's designation scope (30%), or the application is not complete (25%). Only 14 of 52 notified bodies reported that more than half of the submissions they receive are complete.
Timelines are long even for complete files. For a new QMS plus product certificate, 51% of notified bodies report 13 to 18 months from signed agreement to certificate, and a further 31% report 19 to 24 months.
Clinical evaluation is where an incomplete file tends to cost the most time, because missing clinical data cannot be produced on request. In files Qserve has supported, insufficient data has pushed reviews beyond 18 months. In those cases the practical question becomes how to close the gap quickly and affordably, which is where PMCF surveys and studies and a realistic reading of imperfect data come in.
How many Notified Body question rounds does a CER review take?
In Qserve's review of the clinical evaluations it wrote or supported, most CERs completed Notified Body clinical review within two rounds of questions. Clearing in a single round was the exception: 24% of class III files and 16% of class IIa files, and none of the class IIb active device or software files.
|
Device class |
Cleared in one round |
Cleared in two rounds |
|
Class III |
24% |
76% |
|
Class IIb (active devices and software) |
0% |
100% |
|
Class IIa |
16% |
83% |
Based on applications where Qserve wrote the clinical evaluation, supported specific parts of the clinical data analysis, or answered Notified Body questions.
The planning lesson is simple. Build the second round into your certification timeline as the expected case, and treat a one-round outcome as time you gain rather than time you were owed.
What decides whether a CER clears review in one round?
Whether a clinical evaluation clears Notified Body review in one round depends on the quality of the clinical data and on whether that data covers every device variant and accessory in scope. Claims that outrun the data also draw questions: in 45% of the files Qserve reviewed, the device lifetime claim had to be adjusted to match the clinical data already available.
Lifetime is a good example of how to handle that gap. Rather than accept a shorter claim by default, Qserve searched the literature for suitable data to extend the lifetime claim as far as the evidence would support it. Variant coverage works the same way. If the technical documentation lists several sizes, configurations, or accessories, the reviewer will ask which data supports each one, so the CER should answer that before the question is asked.
Why does endpoint selection matter so much in clinical evaluation?
Clinical endpoints, also called clinical outcome parameters, determine what a clinical evaluation can and cannot demonstrate about a device's safety and performance. Qserve treats endpoint selection as the most important element of a clinical evaluation, because Notified Bodies compare complaint rates and endpoints against similar devices and use that comparison to direct their in-depth questions.
The practical fix is to choose endpoints from the literature that match the endpoints used in your post-market clinical follow-up activities. When the literature, the PMCF data, and the CER measure the same things, the analysis is straightforward and the comparison a reviewer runs works in your favor.
What if a Notified Body reviewer applies a personal preference?
In rare cases, a Notified Body clinical expert gives more weight to negative literature on a device than the full body of evidence supports, typically where several treatment options exist and the reviewer favors one of them. Some Notified Bodies brief their clinical experts on what is sufficient from a regulatory perspective, which removes most of this risk.
Where that briefing has not happened, the answer is to bring the discussion back to the regulatory requirement: what the MDR and the relevant guidance ask the manufacturer to demonstrate, and where the evidence meets that bar. Qserve has successfully argued these points with Notified Bodies when a finding rested on preference rather than on the regulation.
Which CER weaknesses draw the most Notified Body questions?
Recurring weaknesses that draw Notified Body questions include a thin state of the art analysis, equivalence claims without a full technical, biological, and clinical comparison, and a CER that is disconnected from post-market data. Jorn van Binsbergen covers the five most common findings in detail in Top 5 Notified Body Findings in CER Reviews.
- State of the art: the benchmark your benefit-risk conclusion is judged against. See what state of the art means in practice.
- Equivalence: for class III and implantable devices, Article 61(5) of the MDR requires a contract giving full access to the equivalent device's technical documentation. See how to use the equivalence pathway under MDR and equivalence for implantable and class III devices.
- Post-market data: PMCF objectives should answer the open questions the CER leaves, and the CER should be updated when post-market data changes the picture. See when your CER needs updating.
How Qserve helps manufacturers prepare for and respond to CER reviews
Qserve's clinical evaluation services combine medical writing, regulatory strategy, and an in-house CRO team, so evidence gaps can be closed in the same engagement that identifies them. Support for CER reviews includes:
- A Notified Body-like review of your clinical evaluation before submission
- A clinical evaluation gap assessment structured on the CEAR template in MDCG 2020-13
- An assessment of sufficient clinical evidence based on the device's clinical claims, indications, and intended purpose
- Cost-effective, time-efficient clinical data collection strategies, including PMCF studies and clinical investigations
- Notified Body query responses written to limit the number of question rounds
- Comprehensive systematic literature reviews
- Selection and documentation of clinical outcome parameters for literature data extraction, clinical investigations, or PMCF surveys
- Full clinical evaluation remediation
Next step: check your CER before the Notified Body does
Not sure how your current clinical evaluation would hold up? Start with the free Clinical Evaluation Readiness Check, or schedule a free introductory call with Qserve's clinical team to scope a CER gap analysis, outsourced CER authoring, or support through an active Notified Body review.
Couldn't find your question?
Does MEDDEV 2.7/1 rev. 4 still apply under EU MDR?
MEDDEV 2.7/1 rev. 4 was written for the Medical Devices Directive, so it is not legally binding under EU MDR. Its methodology, such as how to appraise literature and structure a clinical evaluation, is still widely used where it is consistent with the MDR and current MDCG guidance. Where they differ, the MDR and MDCG guidance take precedence.
Is clinical data from an MDD-era CER enough for MDR certification?
Not automatically. MDCG 2020-6 states that clinical data used for conformity assessment under the directives does not necessarily provide sufficient clinical evidence under the MDR. Manufacturers of legacy devices are expected to update the clinical evaluation plan, run a gap analysis against the MDR general safety and performance requirements, and close the gaps they find.
What is a Notified Body-like review?
A Notified Body-like review is an independent assessment of a clinical evaluation before it is submitted, carried out by reviewers who ask the questions a Notified Body would ask. It identifies gaps in evidence, claims, and traceability while they can still be fixed, rather than after they have become formal Notified Body questions.